左側相關訊息
Shodex

CD-200


 - The electric conductivity detector is
    designed for ion chromatography.
 - It is recommended for anion or cation
    analysis in aqueous solution.


CD-200 詳細資訊 *

Features


 - With use of the built-in double temperature control cell,
    this detector enables highly sensitive measurement.
 - The detector supports a wide range of measurement and is usable for
    ion chromatography with or without suppressor method.


Specifications
 

Model

CD-200

Measurement method

Two-electorode system

Measurement limit 0 to 600mS/m (0 to 6mS/cm)
Measurement range 0.0025 to 5.12mS/m, 0.025 to 51.2mS/m, 0.25 to 512mS/m
Linearity range 600mS/m
Response 0.1, 0.25, 0.5, 1.0, 1.5, 2, 3, 6sec
Auto zero limit Same as measurement limit
Baseline shift Range;0 to 2mS/m, Resolution;0.01mS/m
Integrator output 0 to 1V (Sensitivity:200, 20, 2mV/mS/m)
Recorder output 0 to 10mV/FS
Cell Temperature control OFF, 30 to 50°C (1°C step), 77°C Temp. fuse
Communication port USB
Cell volume 2.5μL
Pressure rating 1MPa
Wetted materials Stainless steel 316, Teflon, PEEK
Dimension, Weight W260 x D400 x H150 (mm), ca. 8kg
Power source
Power consumption
AC 100 to 240V±10%, 200VA max

 
-----------------------------------------------------------------------------------------------------------------

**尚有其它系列、規格產品

   如需產品報價或詳細資料,歡迎來電或來信洽詢~~